This standard is recognized by standards organizations such the American National Standards Institute (ANSI) and the European Commission in IEC60601-1 and requires that RACM1200-V provides Means of Operator Protection (MOOP) isolation of 3000Vac, maintain creepage and clearance distances of 5mm and 4mm, and provide double insulation between the primary and secondary ends of the power supply.
Additionally, RACM1200-V meets low leakage current limits for BF-rated (Body Floating) applications, as set by the IEC 60601-1, that are normally conductive and are often attached to a patient’s body for a medium or long term amount of time. These applications include but are not limited to blood pressure monitors, incubators, ultrasound equipment, heart rate monitors, and more. The below table outlines the isolation capabilities of BF-rated devices, which RACM1200-V conforms to and even exceeds.
RACM1200-V supports POE requirements by providing 2000VAC isolation barriers to ground. Lastly, the unit is safety certified to IEC/EN 62368, UL/IEC/EN 60601; IEC/EN 61558-2-16.